Dissolved Oxygen Pen
Meter
850045
Instruction Manual
Vedere la pagina 0
1 2 3 4 5 6 ... 15 16

Sommario

Pagina 1 - Dissolved Oxygen Pen

Dissolved Oxygen Pen Meter 850045 Instruction Manual

Pagina 2 - TABLE OF CONTENTS

PROBE MAINTENANCE Initial Use 1. Unpack the DO meter and fill the probe head with the Electrolyte (OXEL-03) fluid provided. 2. Follow the steps sh

Pagina 3 - FEATURES

11 PROBE MAINTENANCE 1. Unscrew the probe head (Fig., 2-3) 2. Pour out the current Electrolyte fluid from the probe head (unless this is the initial

Pagina 4

Display LCD size: 29mm x 28mm Measurement & Range Dissolved Oxygen: 0 to 20.0 mg/L Temperature: 0 to 50 ºC 32 to 122 ºF R

Pagina 5 -

SPECIFICATIONS 13 Water Resistance IP-65: 0 to 50 ºC - 32 to 122 ºF IP-67: 0 to 30 ºC - 32 to 86 ºF Dimension 180 x 40 x 40mm 8.1” x 1.9” x 1.6” Acce

Pagina 6

1. Replace the batteries when the low battery icon is displayed in the upper left corner of the LCD. Use 4 new AAA batteries. 2. To replace the ba

Pagina 7 - MEASUREMENT PROCEDURES

Sper Scientific warrants this product against defects in materials and workmanship for a period of one (1) years from the date of purchase, and agree

Pagina 9

TABLE OF CONTENTS Features …………………………………...…...……... 3 Front Panel Description ……………………………… 4 Measurement Procedures ………………………..…. 7-9 Probe Mainte

Pagina 10 - PROBE MAINTENANCE

3  Heavy duty polarographic D.O. probe including temperature sensor  Automatic Temperature Compensation  Microprocessor assures high ac

Pagina 11

FRONT PANEL DESCRIPTION 4 FIGURE 1 1-6 1-1 1-7 1-5 1-2 1-3 1-4 1-1 1-2 1-3 1-4 1-5 1-6 1-7

Pagina 12 - SPECIFICATIONS

CALIBRATION 5 Calibration 1. Fill the D.O. meter probe head (completely) with Electrolyte prior to using the first time. Please refer to the

Pagina 13

Perform the calibration procedure in an open and well ventilated area to achieve the best possible results. Once the calibration is complete the m

Pagina 14 - BATTERY REPLACEMENT

MEASUREMENT PROCEDURES 7 Dissolved Oxygen (DO) Measurement 1. Make certain the meter has been calibrated (page 5). 2. Press and hold the Unit butt

Pagina 15 - Revised: 5/1/2012

MEASUREMENT PROCEDURES 8 Temperature Measurement 1. When not in calibration mode, the meter continually displays the temperature at the bottom le

Pagina 16

Data Record (Maximum & Minimum Reading) 1. The data record function records the maximum and min-imum readings. Press the REC button (Fig., 1-4)