Sper-scientific 850037 Large Display Conductivity Pen Manuale Utente Pagina 10

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 9
10
5. Press SET for less than one second to
conrm the selection or press SET for
more than 2 seconds to return to P3.0
without saving.
6. When in P3.0, press ▲ to advance to
the next setup function or press SET
for more than 2 seconds to return to
Measurement Mode.
Calibration Review (CAL)
1. Press SET for 2 seconds to enter setup.
2. Press ▲ to select the calibration review
setting. “CAL” will appear on the primary
display of the LCD and “P4.0” will
appear on the secondary display.
Range 1 Calibration Information:
3. When in P4.0, press SET for less than
one second to enter P4.1 to review the
last calibration concentration. If the pen
is not yet calibrated, “---” will appear on
the LCD.
Vedere la pagina 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 55 56

Commenti su questo manuale

Nessun commento