5. CALIBRATION PROCEDURE
1)The meter can be calibrated under following calibration
points :
Free chlorine
*Zero
*1.00 ppm
or
Total chlorine
*Zero
*1.00 ppm
2)The meter ship along with
* Zero Chlorine standard solution ( CL-01 ) X 1 PC,
* 1.0 ppm Free Chlorine standard solution ( CF-01 ) X 1 PC.
* 1.0 ppm Total Chlorine standard solution ( CT-01) X 1 PC.
as the standard accessories.
3)The complete calibration should be executed by
following two solution :
*Zero Chlorine standard solution
*1.0 ppm standard solution ( Free or Total ).